Dr.Sarah Qadeer
Introduction: Since the introduction of osseointegrated dental implants, long-term clinical studies have confirmed the efficiency of implant therapy.1-2. Despite this, there are many kinds of complications associated with dental implant therapy, which may range from screw loosening, occlusal material failure, prosthesis fractures, implant fractures and in severe cases, peri-implant bone loss.
Excessive loading on the peri-implant bone is one of the major causes for implant failure and occlusal material failure. Published studies suggest that, due to the absence of the periodontal ligament around an implant with prosthesis upon it, a lack of normal proprioreception results in the patients exhibiting the inability to sense the degree of force delivered occlusally to the prosthesis during normal function. Excessive occlusal loading on dental implants has been implicated in the failure of both material and implant osseointegration, although a controversy exists as to the role of occlusion and bone loss following implant placement, but there is a generalized agreement that early implant failure may be associated with overload.3.
Concept and Rationale for Delayed Loading on Implant Prosthesis:
The fundamental, inherent difference between the tooth and implant is that an endosseous implant is in direct contact with the bone while a natural tooth is suspended by perio-dontal ligament which acts as shock absorber for excessive occlusal forces.
Causes of implant failure can be categorized as 1) defects in the design of the material, 2) non-passive fit of prosthetic structure, and 3) biomechanical or physiologic overload 4 . Chewing mainly involves vertical forces but horizontal movement of the mandible and inclination of the dental cusps cause lateral forces that are transferred to implant and the bone. Due to the lack of periodontal ligaments around the implant prosthesis, the depressability of implant prosthesis is five times less than the normal teeth,5 and patients are 8 times less able to discern tactilely occlusal force excess6, such that a desirable occlusal endpoint goal is to occlude the natural teeth prior to the implant prosthesis, so as to minimize excessive load on the implant prosthesis.7
The goal of implant occlusion have been described by Chapman and colleagues as bilateral simultaneous contact, freedom from protrusive and lateral interferences, absence of non-working contact and equal distribution of occlusal forces8. It is also recognized that excessive loading or incorrectly vectored forces as well as time from implant placement to loading can place deleterious stresses on the implant resulting in failure.9
Use of T-Scan Computerized Occlusal Analysis to achieve Time Delay on Implant Prosthesis
Maness described the concept of relative occlusal force which advocates equilibrating the distribution of forces on teeth capable of bearing occlusal loads by using computerized occlusal analysis provided by the T-Scan computer (Tekscan Inc. Boston)10. T-Scan computerized occlusal recording can give a precise time over force analysis and pinpoint the irregularities in terms of percentage of occlusal force distribution on individual tooth to help explore and analyze the patient’s occlusion that can be visualized as dynamic images to provide necessary maintenance post implant delivery.
Since the implant retained prosthesis exhibits less depressibility than the natural teeth, simultaneous occlusal loading on the natural as well as the implant teeth within the same quadrant may result in implant prosthesis bearing more of the occlusal load than the more depressible natural teeth. Kerstein proposed a quantifiable time delay (<.4 SECONDS) such that the natural teeth occlude prior to the implant prosthesis8. The natural teeth would undergo force dissipation within the periodontal ligament sling, after which occlusal loading on the Implant prosthesis would occur. This time delay must be short enough so that the Implant Prosthesis actually occludes.
Case 1: Implant Prosthesis placed on Left maxillary Arch, tooth no. 15, 16 &17 of a 57 year old female patient.
Fig.1a. Initial Recording: Screen capture of T-Scan recording. Implant Prosthesis on tooth # 15, 16 & 17 represented in red in the arch showing 73% force on Implant side.
Fig.1b. Post adjustment: Natural teeth come into contact first followed by implant prosthesis at 0.554 seconds.

Case 2: Maxillary Implants placed in the anterior region of a 42 year old male patient at Gangdong KyungHee University Dental Hospital in S.Korea. Images provided by Prof. Richard Lee Sung-Bok.
Case 2 Fig. 2d. The natural teeth undergo force dissipation within the periodontal ligament sling, after which occlusal loading on the Implant prosthesis occurs at 0.45 seconds with well balanced forces on either sides (55.9% on left side Vs 44.1% on right side). This brief time delay prevents excessive loading on the implant prosthesis.
Conclusion: The goal of a dentist is to develop a balanced and harmonious occlusion following dental treatment. In order to do so, digital technology can improve prosthesis outcome by providing precise measurable data in terms of force, timing and sequence of closure to make accurate adjustments in the occlusion. This quantifiable data is available in the form of a comprehensive movie that can be analysed and balancing can be achieved to satisfaction.
T-Scan technology provides comprehensive visual insight to the occlusion for the clinician to explore as well as the patient to engage their attention for successful treatment outcome.
By bringing about delayed loading on implant prosthesis (by <0.4 seconds) enhances implant longetivity as the detrimental excessive occlusal force can be monitored and minimized using T-Scan computerized occlusal analysis, thus preventing occlusal overload on the implant prosthesis.
References:
1. Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;10:387-416.
2. Albrektsson T, Zarb G, Worthington P, Eriksson AR. The long-term efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac. 1986; 1:11-25
3. Esposito M, Hirsch JM, Lekholm U, et al. Biological factors contributing to failures of osseointegrated oral implants. Eur J Oral Sci.1998; 106:721-764.
4. Balshi TJ. An analysis and management of fractured implants: a clinical report. Int J Oral Maxillofac Implants. 1996; 11:660-666
5. Sekine H. et al: Mobility characteristics and tactile sensitivity of osseointegrated fixture- supporting systems, in Van Steenberghe D, editor: Tissue integration in oral maxillofacial reconstruction, Excerpta Medica, Amsterdam, Elsevier, 1986; pp.306-332.
6. Hammerle CH, Wagner D, Bragger U, Lussi A, Karayiannis A, Joss A, Lang NP: Threshold of tactile sensitivity perceived with dental endosseous implants and natural teeth. Clin Oral Implants Res. 1995 Jun; 6(2):83-90.
7. Kerstein, R.B., Nonsimultaneous Tooth Contact in Combined Implant and Natural Tooth Occlusal Schemes, PPAD, 2002 Vol. 13, No. 9, pp. 751-756..
8. Chapman R. Principles of occlusion for implant prostheses: guidelines for position, timing, and force of occlusal contacts. Quintessence Int.1987; 20:473-80
9. Christopher J.Stevens. Computerized Occlusal Implant management with the T-Scan II system: A case report: Dentistry Today 2006 Vol. 25 No.2 pp. 1-4
10. Maness W. T-Scan Clinical applications manual: Tekscan Inc Boston; 1988